| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.300 | 2.217 |
| 1000.00 | 1.487 | 1.404 |
| 500.00 | 1.165 | 1.082 |
| 250.00 | 0.859 | 0.776 |
| 125.00 | 0.517 | 0.434 |
| 62.50 | 0.327 | 0.244 |
| 31.25 | 0.285 | 0.202 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-110% | 99% |
| EDTA plasma(n=5) | 82-122% | 102% |
| Heparin plasma(n=5) | 90-109% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-96% | 93-114% | 93-111% | 81-100% |
| EDTA plasma(n=5) | 86-105% | 87-120% | 82-91% | 94-115% |
| Heparin plasma(n=5) | 95-106% | 95-120% | 81-106% | 84-115% |