| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.188 | 2.083 |
| 500.00 | 1.549 | 1.444 |
| 250.00 | 1.232 | 1.127 |
| 125.00 | 0.766 | 0.661 |
| 62.50 | 0.601 | 0.496 |
| 31.25 | 0.385 | 0.280 |
| 15.63 | 0.203 | 0.098 |
| 0.00 | 0.105 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-116% | 101% |
| EDTA plasma(n=5) | 92-105% | 99% |
| Heparin plasma(n=5) | 81-102% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-87% | 87-111% | 95-102% | 95-110% |
| EDTA plasma(n=5) | 84-115% | 86-119% | 91-100% | 83-99% |
| Heparin plasma(n=5) | 84-120% | 84-113% | 89-95% | 91-99% |