| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.018 | 1.929 |
| 5.00 | 1.655 | 1.566 |
| 2.50 | 1.133 | 1.044 |
| 1.25 | 0.875 | 0.786 |
| 0.63 | 0.518 | 0.429 |
| 0.32 | 0.369 | 0.280 |
| 0.16 | 0.196 | 0.107 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-97% | 90% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 89-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-96% | 82-95% | 81-93% |
| EDTA plasma(n=5) | 85-92% | 79-96% | 95-102% | 87-98% |
| Heparin plasma(n=5) | 79-91% | 96-105% | 80-92% | 91-103% |