| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.086 | 1.991 |
| 5.00 | 1.750 | 1.655 |
| 2.50 | 1.231 | 1.136 |
| 1.25 | 0.833 | 0.738 |
| 0.63 | 0.519 | 0.424 |
| 0.31 | 0.306 | 0.211 |
| 0.16 | 0.208 | 0.113 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-113% | 103% |
| EDTA plasma(n=5) | 89-126% | 108% |
| Heparin plasma(n=5) | 83-105% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-95% | 82-100% | 91-105% | 94-107% |
| EDTA plasma(n=5) | 88-111% | 80-103% | 95-119% | 81-103% |
| Heparin plasma(n=5) | 83-117% | 84-112% | 94-107% | 81-116% |