| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.214 | 2.128 |
| 2500.00 | 1.593 | 1.507 |
| 1250.00 | 1.106 | 1.020 |
| 625.00 | 0.697 | 0.611 |
| 312.50 | 0.519 | 0.433 |
| 156.25 | 0.357 | 0.271 |
| 78.13 | 0.226 | 0.140 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 97-105% | 101% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 90-105% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-102% | 93-106% | 87-101% | 91-104% |
| EDTA plasma(n=5) | 89-97% | 93-102% | 81-96% | 93-106% |
| Heparin plasma(n=5) | 86-93% | 91-105% | 87-98% | 89-94% |