| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.948 | 1.852 |
| 5.00 | 1.581 | 1.485 |
| 2.50 | 1.173 | 1.077 |
| 1.25 | 0.900 | 0.804 |
| 0.63 | 0.579 | 0.483 |
| 0.31 | 0.328 | 0.232 |
| 0.16 | 0.176 | 0.080 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-99% | 96% |
| EDTA plasma(n=5) | 94-128% | 111% |
| Heparin plasma(n=5) | 93-115% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-109% | 80-113% | 92-108% | 84-100% |
| EDTA plasma(n=5) | 81-86% | 85-96% | 87-105% | 95-116% |
| Heparin plasma(n=5) | 90-119% | 81-99% | 84-111% | 94-110% |