| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.316 | 2.227 |
| 10.00 | 1.644 | 1.555 |
| 5.00 | 0.992 | 0.903 |
| 2.50 | 0.780 | 0.691 |
| 1.25 | 0.583 | 0.494 |
| 0.63 | 0.350 | 0.261 |
| 0.31 | 0.226 | 0.137 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-111% | 102% |
| EDTA plasma(n=5) | 81-100% | 91% |
| Heparin plasma(n=5) | 81-106% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-110% | 88-100% | 81-115% | 87-95% |
| EDTA plasma(n=5) | 91-105% | 90-97% | 91-111% | 92-106% |
| Heparin plasma(n=5) | 86-103% | 91-115% | 84-120% | 83-114% |