| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.126 | 2.043 |
| 5.00 | 1.527 | 1.444 |
| 2.50 | 1.183 | 1.100 |
| 1.25 | 0.975 | 0.892 |
| 0.63 | 0.543 | 0.460 |
| 0.32 | 0.353 | 0.270 |
| 0.16 | 0.196 | 0.113 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-99% | 92% |
| EDTA plasma(n=5) | 96-107% | 102% |
| Heparin plasma(n=5) | 83-95% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-107% | 83-99% | 80-96% | 91-101% |
| EDTA plasma(n=5) | 85-97% | 86-102% | 87-99% | 99-105% |
| Heparin plasma(n=5) | 95-104% | 82-94% | 86-96% | 81-97% |