| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.126 | 2.037 |
| 50.00 | 1.638 | 1.549 |
| 25.00 | 0.935 | 0.846 |
| 12.50 | 0.828 | 0.739 |
| 6.25 | 0.572 | 0.483 |
| 3.13 | 0.379 | 0.290 |
| 1.57 | 0.258 | 0.169 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-93% | 85% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 90-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |
| EDTA plasma(n=5) | 85-94% | 87-101% | 89-99% | 85-95% |
| Heparin plasma(n=5) | 86-98% | 93-106% | 89-97% | 85-94% |