| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.842 | 1.753 |
| 5.00 | 1.494 | 1.405 |
| 2.50 | 1.088 | 0.999 |
| 1.25 | 0.835 | 0.746 |
| 0.63 | 0.591 | 0.502 |
| 0.32 | 0.339 | 0.250 |
| 0.16 | 0.175 | 0.086 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-92% | 86% |
| EDTA plasma(n=5) | 78-95% | 86% |
| Heparin plasma(n=5) | 80-96% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 86-92% | 95-103% | 89-97% |
| EDTA plasma(n=5) | 79-94% | 87-96% | 82-95% | 89-102% |
| Heparin plasma(n=5) | 85-92% | 87-103% | 90-98% | 95-102% |