| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 1.985 | 1.898 |
| 50.00 | 1.586 | 1.499 |
| 25.00 | 1.170 | 1.083 |
| 12.50 | 0.741 | 0.654 |
| 6.25 | 0.533 | 0.446 |
| 3.13 | 0.333 | 0.246 |
| 1.56 | 0.186 | 0.099 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-112% | 103% |
| EDTA plasma(n=5) | 86-103% | 95% |
| Heparin plasma(n=5) | 81-94% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-90% | 81-104% | 91-110% | 86-114% |
| EDTA plasma(n=5) | 92-117% | 83-106% | 85-112% | 90-99% |
| Heparin plasma(n=5) | 87-111% | 88-108% | 85-96% | 82-98% |