| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 300.00 | 1.919 | 1.815 |
| 150.00 | 1.583 | 1.479 |
| 75.00 | 1.128 | 1.024 |
| 37.50 | 0.828 | 0.724 |
| 18.75 | 0.453 | 0.349 |
| 9.38 | 0.324 | 0.220 |
| 4.69 | 0.192 | 0.088 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 83-99% | 95% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-100% | 93-108% | 80-101% | 87-96% |
| EDTA plasma(n=5) | 86-102% | 87-96% | 89-98% | 83-103% |
| Heparin plasma(n=5) | 85-97% | 87-96% | 85-96% | 85-95% |