| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.952 | 1.865 |
| 5.00 | 1.790 | 1.703 |
| 2.50 | 1.182 | 1.095 |
| 1.25 | 0.855 | 0.768 |
| 0.63 | 0.573 | 0.486 |
| 0.32 | 0.313 | 0.226 |
| 0.16 | 0.256 | 0.169 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-95% | 86% |
| EDTA plasma(n=5) | 80-96% | 88% |
| Heparin plasma(n=5) | 80-92% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-102% | 86-102% | 87-99% | 99-105% |
| EDTA plasma(n=5) | 85-98% | 88-104% | 86-102% | 79-93% |
| Heparin plasma(n=5) | 98-105% | 92-101% | 85-105% | 87-98% |