| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 20.00 | 2.266 | 2.171 | 
| 10.00 | 1.783 | 1.688 | 
| 5.00 | 1.165 | 1.070 | 
| 2.50 | 0.971 | 0.876 | 
| 1.25 | 0.547 | 0.452 | 
| 0.63 | 0.335 | 0.240 | 
| 0.32 | 0.179 | 0.084 | 
| 0.00 | 0.095 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 92-105% | 98% | 
| EDTA plasma(n=5) | 84-96% | 90% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-96% | 92-105% | 89-102% | 
| EDTA plasma(n=5) | 85-102% | 79-92% | 87-103% | 89-98% | 
| Heparin plasma(n=5) | 79-97% | 96-104% | 85-93% | 89-97% |