| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 20.00 | 1.946 | 1.852 | 
| 10.00 | 1.519 | 1.425 | 
| 5.00 | 1.175 | 1.081 | 
| 2.50 | 0.933 | 0.839 | 
| 1.25 | 0.567 | 0.473 | 
| 0.63 | 0.388 | 0.294 | 
| 0.32 | 0.215 | 0.121 | 
| 0.00 | 0.094 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-95% | 87% | 
| EDTA plasma(n=5) | 83-97% | 90% | 
| Heparin plasma(n=5) | 78-92% | 85% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-104% | 87-96% | 89-101% | 97-105% | 
| EDTA plasma(n=5) | 89-102% | 79-94% | 83-96% | 85-106% | 
| Heparin plasma(n=5) | 98-105% | 96-103% | 87-95% | 83-92% |