 
| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.285 | 2.203 | 
| 5.00 | 1.711 | 1.629 | 
| 2.50 | 1.226 | 1.144 | 
| 1.25 | 0.867 | 0.785 | 
| 0.63 | 0.543 | 0.461 | 
| 0.32 | 0.364 | 0.282 | 
| 0.16 | 0.134 | 0.052 | 
| 0.00 | 0.082 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 90-104% | 97% | 
| Heparin plasma(n=5) | 80-97% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-105% | 84-98% | 95-102% | 88-102% | 
| EDTA plasma(n=5) | 81-92% | 92-103% | 87-98% | 90-101% | 
| Heparin plasma(n=5) | 89-99% | 83-103% | 98-105% | 96-101% |