| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.345 | 2.254 | 
| 2500.00 | 1.777 | 1.686 | 
| 1250.00 | 1.182 | 1.091 | 
| 625.00 | 0.864 | 0.773 | 
| 312.50 | 0.555 | 0.464 | 
| 156.25 | 0.239 | 0.148 | 
| 78.13 | 0.206 | 0.115 | 
| 0.00 | 0.091 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 94-105% | 98% | 
| EDTA plasma(n=5) | 82-94% | 88% | 
| Heparin plasma(n=5) | 83-99% | 95% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-95% | 83-96% | 92-102% | 82-93% | 
| EDTA plasma(n=5) | 83-97% | 92-103% | 87-98% | 93-101% | 
| Heparin plasma(n=5) | 85-96% | 86-95% | 90-102% | 81-90% |