| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.160 | 2.057 |
| 5.00 | 1.554 | 1.451 |
| 2.50 | 1.224 | 1.121 |
| 1.25 | 0.806 | 0.703 |
| 0.63 | 0.496 | 0.393 |
| 0.31 | 0.388 | 0.285 |
| 0.16 | 0.228 | 0.125 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-103% | 93% |
| EDTA plasma(n=5) | 90-121% | 106% |
| Heparin plasma(n=5) | 87-113% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-114% | 94-104% | 85-106% | 92-119% |
| EDTA plasma(n=5) | 95-100% | 90-100% | 93-108% | 89-96% |
| Heparin plasma(n=5) | 95-113% | 87-107% | 81-93% | 81-96% |