| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.035 | 1.949 | 
| 5.00 | 1.576 | 1.490 | 
| 2.50 | 1.107 | 1.021 | 
| 1.25 | 0.915 | 0.829 | 
| 0.63 | 0.621 | 0.535 | 
| 0.32 | 0.379 | 0.293 | 
| 0.16 | 0.238 | 0.152 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-95% | 87% | 
| EDTA plasma(n=5) | 84-98% | 91% | 
| Heparin plasma(n=5) | 79-97% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-103% | 85-93% | 89-97% | 95-102% | 
| EDTA plasma(n=5) | 83-95% | 85-96% | 87-101% | 83-95% | 
| Heparin plasma(n=5) | 97-105% | 88-95% | 96-102% | 83-97% |