| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.108 | 2.020 | 
| 5.00 | 1.674 | 1.586 | 
| 2.50 | 1.106 | 1.018 | 
| 1.25 | 0.847 | 0.759 | 
| 0.63 | 0.593 | 0.505 | 
| 0.32 | 0.388 | 0.300 | 
| 0.16 | 0.196 | 0.108 | 
| 0.00 | 0.088 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 90-103% | 96% | 
| EDTA plasma(n=5) | 81-96% | 88% | 
| Heparin plasma(n=5) | 78-94% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-104% | 97-105% | 82-95% | 89-97% | 
| EDTA plasma(n=5) | 82-95% | 89-103% | 79-91% | 85-93% | 
| Heparin plasma(n=5) | 78-91% | 86-92% | 89-96% | 91-97% |