| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 1.958 | 1.872 | 
| 500.00 | 1.638 | 1.552 | 
| 250.00 | 1.277 | 1.191 | 
| 125.00 | 0.937 | 0.851 | 
| 62.50 | 0.538 | 0.452 | 
| 31.25 | 0.381 | 0.295 | 
| 15.63 | 0.168 | 0.082 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 91-105% | 98% | 
| EDTA plasma(n=5) | 85-97% | 91% | 
| Heparin plasma(n=5) | 80-94% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 82-103% | 81-91% | 85-101% | 81-90% | 
| EDTA plasma(n=5) | 83-92% | 92-96% | 86-98% | 95-101% | 
| Heparin plasma(n=5) | 89-101% | 79-92% | 82-93% | 83-97% |