| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.115 | 2.048 |
| 1000.00 | 1.567 | 1.500 |
| 500.00 | 1.275 | 1.208 |
| 250.00 | 0.921 | 0.854 |
| 125.00 | 0.607 | 0.540 |
| 62.50 | 0.386 | 0.319 |
| 31.25 | 0.224 | 0.157 |
| 0.00 | 0.067 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-107% | 95% |
| EDTA plasma(n=5) | 94-115% | 105% |
| Heparin plasma(n=5) | 94-105% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-119% | 89-106% | 91-113% | 90-102% |
| EDTA plasma(n=5) | 89-104% | 92-107% | 90-111% | 93-113% |
| Heparin plasma(n=5) | 94-107% | 81-92% | 86-93% | 91-108% |