| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 2.096 | 2.027 | 
| 1000.00 | 1.612 | 1.543 | 
| 500.00 | 1.283 | 1.214 | 
| 250.00 | 0.898 | 0.829 | 
| 125.00 | 0.601 | 0.532 | 
| 62.50 | 0.379 | 0.310 | 
| 31.25 | 0.225 | 0.156 | 
| 0.00 | 0.069 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 91-99% | 95% | 
| EDTA plasma(n=5) | 80-92% | 86% | 
| Heparin plasma(n=5) | 85-97% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 84-99% | 83-101% | 86-99% | 93-101% | 
| EDTA plasma(n=5) | 89-98% | 81-94% | 96-105% | 92-103% | 
| Heparin plasma(n=5) | 88-97% | 91-105% | 87-98% | 95-105% |