| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.155 | 2.068 | 
| 5.00 | 1.577 | 1.490 | 
| 2.50 | 1.243 | 1.156 | 
| 1.25 | 0.905 | 0.818 | 
| 0.63 | 0.532 | 0.445 | 
| 0.32 | 0.329 | 0.242 | 
| 0.16 | 0.222 | 0.135 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 92-105% | 98% | 
| EDTA plasma(n=5) | 95-105% | 100% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 81-94% | 93-107% | 93-102% | 91-105% | 
| EDTA plasma(n=5) | 86-105% | 84-98% | 95-102% | 88-106% | 
| Heparin plasma(n=5) | 87-94% | 91-104% | 95-107% | 82-96% |