| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.164 | 2.085 | 
| 2500.00 | 1.588 | 1.509 | 
| 1250.00 | 1.185 | 1.106 | 
| 625.00 | 0.777 | 0.698 | 
| 312.50 | 0.627 | 0.548 | 
| 156.25 | 0.325 | 0.246 | 
| 78.13 | 0.212 | 0.133 | 
| 0.00 | 0.079 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-95% | 88% | 
| EDTA plasma(n=5) | 83-95% | 89% | 
| Heparin plasma(n=5) | 85-97% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 81-94% | 96-105% | 92-103% | 82-96% | 
| EDTA plasma(n=5) | 89-101% | 87-98% | 92-104% | 82-96% | 
| Heparin plasma(n=5) | 93-102% | 80-93% | 91-100% | 88-95% |