| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.252 | 2.144 | 
| 5.00 | 1.586 | 1.478 | 
| 2.50 | 1.138 | 1.030 | 
| 1.25 | 0.899 | 0.791 | 
| 0.63 | 0.513 | 0.405 | 
| 0.32 | 0.351 | 0.243 | 
| 0.16 | 0.203 | 0.095 | 
| 0.00 | 0.108 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-95% | 91% | 
| EDTA plasma(n=5) | 81-95% | 88% | 
| Heparin plasma(n=5) | 83-97% | 90% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-104% | 93-102% | 79-94% | 87-98% | 
| EDTA plasma(n=5) | 89-99% | 88-104% | 84-98% | 95-105% | 
| Heparin plasma(n=5) | 78-92% | 83-96% | 95-107% | 87-102% |