| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.103 | 2.007 | 
| 5.00 | 1.562 | 1.466 | 
| 2.50 | 1.217 | 1.121 | 
| 1.25 | 0.799 | 0.703 | 
| 0.63 | 0.579 | 0.483 | 
| 0.32 | 0.336 | 0.240 | 
| 0.16 | 0.135 | 0.039 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-95% | 89% | 
| EDTA plasma(n=5) | 88-102% | 95% | 
| Heparin plasma(n=5) | 79-97% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-102% | 98-106% | 87-98% | 91-102% | 
| EDTA plasma(n=5) | 89-98% | 81-94% | 96-105% | 92-103% | 
| Heparin plasma(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |