| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 20.00 | 1.954 | 1.877 | 
| 10.00 | 1.573 | 1.496 | 
| 5.00 | 1.085 | 1.008 | 
| 2.50 | 0.861 | 0.784 | 
| 1.25 | 0.457 | 0.380 | 
| 0.63 | 0.334 | 0.257 | 
| 0.32 | 0.174 | 0.097 | 
| 0.00 | 0.077 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-93% | 90% | 
| EDTA plasma(n=5) | 80-90% | 85% | 
| Heparin plasma(n=5) | 84-106% | 100% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-98% | 85-92% | 79-96% | 89-100% | 
| EDTA plasma(n=5) | 91-102% | 89-97% | 86-102% | 94-106% | 
| Heparin plasma(n=5) | 81-97% | 88-102% | 85-93% | 83-97% |