| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.119 | 2.011 | 
| 5.00 | 1.722 | 1.614 | 
| 2.50 | 1.169 | 1.061 | 
| 1.25 | 0.829 | 0.721 | 
| 0.63 | 0.596 | 0.488 | 
| 0.32 | 0.342 | 0.234 | 
| 0.16 | 0.195 | 0.087 | 
| 0.00 | 0.108 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 84-106% | 100% | 
| EDTA plasma(n=5) | 87-93% | 90% | 
| Heparin plasma(n=5) | 80-90% | 85% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-99% | 83-95% | 88-94% | 96-101% | 
| EDTA plasma(n=5) | 85-97% | 90-99% | 82-101% | 93-105% | 
| Heparin plasma(n=5) | 86-93% | 91-104% | 87-98% | 95-99% |