| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 1.898 | 1.803 | 
| 5.00 | 1.608 | 1.513 | 
| 2.50 | 1.147 | 1.052 | 
| 1.25 | 0.705 | 0.610 | 
| 0.63 | 0.469 | 0.374 | 
| 0.32 | 0.372 | 0.277 | 
| 0.16 | 0.208 | 0.113 | 
| 0.00 | 0.095 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-95% | 88% | 
| EDTA plasma(n=5) | 87-99% | 93% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 93-102% | 81-96% | 93-106% | 82-95% | 
| EDTA plasma(n=5) | 88-101% | 90-98% | 92-106% | 85-97% | 
| Heparin plasma(n=5) | 87-98% | 92-101% | 82-98% | 89-97% |