 
| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 1.972 | 1.891 | 
| 1000.00 | 1.630 | 1.549 | 
| 500.00 | 1.088 | 1.007 | 
| 250.00 | 0.857 | 0.776 | 
| 125.00 | 0.479 | 0.398 | 
| 62.50 | 0.389 | 0.308 | 
| 31.25 | 0.212 | 0.131 | 
| 0.00 | 0.081 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-95% | 88% | 
| EDTA plasma(n=5) | 87-99% | 93% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-103% | 83-94% | 92-105% | 89-97% | 
| EDTA plasma(n=5) | 88-102% | 81-96% | 86-93% | 82-95% | 
| Heparin plasma(n=5) | 90-99% | 95-106% | 87-101% | 82-97% |