| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 50.00 | 2.131 | 2.033 | 
| 25.00 | 1.563 | 1.465 | 
| 12.50 | 1.196 | 1.098 | 
| 6.25 | 0.896 | 0.798 | 
| 3.13 | 0.473 | 0.375 | 
| 1.57 | 0.413 | 0.315 | 
| 0.79 | 0.232 | 0.134 | 
| 0.00 | 0.098 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 94-107% | 100% | 
| EDTA plasma(n=5) | 82-95% | 88% | 
| Heparin plasma(n=5) | 95-107% | 101% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 82-95% | 88-102% | 84-97% | 89-107% | 
| EDTA plasma(n=5) | 82-94% | 85-96% | 78-90% | 91-99% | 
| Heparin plasma(n=5) | 85-97% | 90-99% | 82-101% | 93-105% |