| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 20.00 | 2.035 | 1.946 | 
| 10.00 | 1.587 | 1.498 | 
| 5.00 | 1.027 | 0.938 | 
| 2.50 | 0.864 | 0.775 | 
| 1.25 | 0.630 | 0.541 | 
| 0.63 | 0.339 | 0.250 | 
| 0.32 | 0.174 | 0.085 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-93% | 87% | 
| EDTA plasma(n=5) | 83-101% | 87% | 
| Heparin plasma(n=5) | 86-94% | 90% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-102% | 86-97% | 89-97% | 79-95% | 
| EDTA plasma(n=5) | 93-102% | 81-96% | 93-106% | 82-95% | 
| Heparin plasma(n=5) | 78-92% | 83-96% | 95-107% | 87-102% |