| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 2.266 | 2.200 | 
| 1000.00 | 1.608 | 1.542 | 
| 500.00 | 1.116 | 1.050 | 
| 250.00 | 0.826 | 0.760 | 
| 125.00 | 0.421 | 0.355 | 
| 62.50 | 0.319 | 0.253 | 
| 31.25 | 0.230 | 0.164 | 
| 0.00 | 0.066 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-93% | 86% | 
| EDTA plasma(n=5) | 94-105% | 98% | 
| Heparin plasma(n=5) | 95-107% | 101% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 91-102% | 86-102% | 87-99% | 99-105% | 
| EDTA plasma(n=5) | 86-95% | 93-103% | 85-94% | 87-101% | 
| Heparin plasma(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |