| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 50.00 | 2.027 | 1.944 | 
| 25.00 | 1.653 | 1.570 | 
| 12.50 | 1.151 | 1.068 | 
| 6.25 | 0.834 | 0.751 | 
| 3.13 | 0.553 | 0.470 | 
| 1.57 | 0.361 | 0.278 | 
| 0.79 | 0.258 | 0.175 | 
| 0.00 | 0.083 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 84-92% | 87% | 
| EDTA plasma(n=5) | 95-104% | 98% | 
| Heparin plasma(n=5) | 82-90% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-92% | 79-96% | 89-100% | 85-96% | 
| EDTA plasma(n=5) | 86-93% | 91-101% | 87-98% | 95-104% | 
| Heparin plasma(n=5) | 98-105% | 92-101% | 85-105% | 87-98% |