| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.179 | 2.085 | 
| 5.00 | 1.617 | 1.523 | 
| 2.50 | 1.226 | 1.132 | 
| 1.25 | 0.847 | 0.753 | 
| 0.63 | 0.511 | 0.417 | 
| 0.32 | 0.324 | 0.230 | 
| 0.16 | 0.113 | 0.019 | 
| 0.00 | 0.094 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 86-99% | 92% | 
| EDTA plasma(n=5) | 80-95% | 87% | 
| Heparin plasma(n=5) | 83-98% | 90% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 93-104% | 83-96% | 95-103% | 88-95% | 
| EDTA plasma(n=5) | 86-97% | 93-105% | 82-93% | 95-104% | 
| Heparin plasma(n=5) | 87-96% | 85-97% | 85-94% | 87-96% |