| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 2.054 | 1.972 | 
| 50.00 | 1.647 | 1.565 | 
| 25.00 | 1.115 | 1.033 | 
| 12.50 | 0.796 | 0.714 | 
| 6.25 | 0.614 | 0.532 | 
| 3.13 | 0.319 | 0.237 | 
| 1.57 | 0.230 | 0.148 | 
| 0.00 | 0.082 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 89-103% | 96% | 
| EDTA plasma(n=5) | 83-97% | 90% | 
| Heparin plasma(n=5) | 86-99% | 92% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 79-93% | 89-97% | 85-94% | 87-101% | 
| EDTA plasma(n=5) | 86-94% | 93-102% | 87-96% | 92-105% | 
| Heparin plasma(n=5) | 91-99% | 88-97% | 88-104% | 97-108% |