Product name: Mouse sIL-6R(SolubleInterleukin 6 Receptor) Microsample ELISA Kit
Reactivity: Mouse
Alternative Names: IL-6R;CD126; gp80; IL6R; IL-6R-1; IL-6R-Alpha; IL6RA; IL-6 receptor subunit alpha; Membrane glycoprotein 80; IL6R; Interleukin 6 Receptor
Assay Type: Sandwich
Sensitivity: 0.118 ng/mL
Standard: 20 ng/mL
Detection Range: 0.32-20 ng/mL
Sample Type: serum, plasma, tissue homogenates and other biological fluids
Assay Length: 3.5h
Research Area: Signal transduction;CD & Adhesion molecule;Infection immunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse sIL-6R. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse sIL-6R. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse sIL-6R, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse sIL-6R in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
20.00 2.194 2.106
10.00 1.579 1.491
5.00 1.224 1.136
2.50 0.908 0.820
1.25 0.526 0.438
0.63 0.372 0.284
0.31 0.220 0.132
0.00 0.088 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant sIL-6R and the recovery rates were calculated by comparing the measured value to the expected amount of sIL-6R in samples.
Matrix Recovery range Average
serum(n=5) 89-116% 103%
EDTA plasma(n=5) 83-125% 104%
Heparin plasma(n=5) 81-100% 91%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of sIL-6R and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 93-111% 86-96% 95-111% 81-88%
EDTA plasma(n=5) 93-120% 84-105% 89-97% 84-97%
Heparin plasma(n=5) 90-107% 87-92% 95-109% 87-119%