| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.379 | 2.297 | 
| 5.00 | 1.517 | 1.435 | 
| 2.50 | 1.182 | 1.100 | 
| 1.25 | 0.864 | 0.782 | 
| 0.63 | 0.555 | 0.473 | 
| 0.32 | 0.324 | 0.242 | 
| 0.16 | 0.227 | 0.145 | 
| 0.00 | 0.082 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-95% | 88% | 
| EDTA plasma(n=5) | 80-94% | 87% | 
| Heparin plasma(n=5) | 89-101% | 95% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-96% | 85-97% | 85-94% | 87-96% | 
| EDTA plasma(n=5) | 85-97% | 87-96% | 85-96% | 85-95% | 
| Heparin plasma(n=5) | 86-96% | 93-104% | 89-97% | 85-94% |