| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 30.00 | 2.095 | 2.006 | 
| 15.00 | 1.841 | 1.752 | 
| 7.50 | 1.034 | 0.945 | 
| 3.75 | 0.887 | 0.798 | 
| 1.88 | 0.556 | 0.467 | 
| 0.94 | 0.351 | 0.262 | 
| 0.47 | 0.156 | 0.067 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 88-103% | 95% | 
| EDTA plasma(n=5) | 95-107% | 101% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 91-99% | 88-97% | 88-104% | 97-108% | 
| EDTA plasma(n=5) | 95-103% | 87-98% | 85-92% | 79-96% | 
| Heparin plasma(n=5) | 86-93% | 93-101% | 86-97% | 90-101% |