| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 1.998 | 1.913 | 
| 5.00 | 1.579 | 1.494 | 
| 2.50 | 1.093 | 1.008 | 
| 1.25 | 0.764 | 0.679 | 
| 0.63 | 0.572 | 0.487 | 
| 0.32 | 0.387 | 0.302 | 
| 0.16 | 0.196 | 0.111 | 
| 0.00 | 0.085 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 84-98% | 91% | 
| EDTA plasma(n=5) | 80-96% | 88% | 
| Heparin plasma(n=5) | 82-98% | 90% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-96% | 82-95% | 93-102% | 87-96% | 
| EDTA plasma(n=5) | 93-106% | 87-99% | 85-93% | 79-96% | 
| Heparin plasma(n=5) | 88-99% | 81-93% | 89-97% | 85-94% |