 
| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 2.045 | 1.964 | 
| 50.00 | 1.593 | 1.512 | 
| 25.00 | 1.186 | 1.105 | 
| 12.50 | 0.728 | 0.647 | 
| 6.25 | 0.563 | 0.482 | 
| 3.13 | 0.426 | 0.345 | 
| 1.57 | 0.288 | 0.207 | 
| 0.00 | 0.081 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 88-102% | 95% | 
| EDTA plasma(n=5) | 95-107% | 101% | 
| Heparin plasma(n=5) | 82-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-103% | 86-92% | 89-97% | 91-103% | 
| EDTA plasma(n=5) | 85-94% | 79-96% | 92-101% | 82-90% | 
| Heparin plasma(n=5) | 85-92% | 79-96% | 92-101% | 87-98% |