| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 4000.00 | 2.268 | 2.172 | 
| 2000.00 | 1.387 | 1.291 | 
| 1000.00 | 0.997 | 0.901 | 
| 500.00 | 0.726 | 0.630 | 
| 250.00 | 0.475 | 0.379 | 
| 125.00 | 0.312 | 0.216 | 
| 62.50 | 0.206 | 0.110 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-95% | 88% | 
| EDTA plasma(n=5) | 90-103% | 96% | 
| Heparin plasma(n=5) | 82-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-104% | 86-102% | 79-93% | 89-97% | 
| EDTA plasma(n=5) | 87-98% | 85-96% | 92-102% | 81-93% | 
| Heparin plasma(n=5) | 95-103% | 79-96% | 93-106% | 87-98% |