| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 8000.00 | 2.053 | 1.962 |
| 4000.00 | 1.648 | 1.557 |
| 2000.00 | 1.036 | 0.945 |
| 1000.00 | 0.761 | 0.670 |
| 500.00 | 0.530 | 0.439 |
| 250.00 | 0.357 | 0.266 |
| 125.00 | 0.233 | 0.142 |
| 0.00 | 0.091 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-94% | 89% |
| EDTA plasma(n=5) | 83-89% | 86% |
| Heparin plasma(n=5) | 92-101% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-99% | 87-109% | 92-105% | 89-111% |
| EDTA plasma(n=5) | 90-102% | 93-100% | 88-104% | 81-101% |
| Heparin plasma(n=5) | 86-105% | 88-102% | 87-104% | 90-117% |