 
| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 20.00 | 2.204 | 2.096 | 
| 10.00 | 1.636 | 1.528 | 
| 5.00 | 1.069 | 0.961 | 
| 2.50 | 0.872 | 0.764 | 
| 1.25 | 0.593 | 0.485 | 
| 0.63 | 0.312 | 0.204 | 
| 0.32 | 0.256 | 0.148 | 
| 0.00 | 0.108 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 87-93% | 90% | 
| Heparin plasma(n=5) | 84-106% | 100% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-97% | 91-103% | 86-92% | 89-97% | 
| EDTA plasma(n=5) | 85-94% | 87-96% | 89-101% | 82-90% | 
| Heparin plasma(n=5) | 78-92% | 96-105% | 87-98% | 95-102% |