| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 1.989 | 1.903 | 
| 250.00 | 1.587 | 1.501 | 
| 125.00 | 1.123 | 1.037 | 
| 62.50 | 0.863 | 0.777 | 
| 31.25 | 0.457 | 0.371 | 
| 15.63 | 0.353 | 0.267 | 
| 7.82 | 0.129 | 0.043 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 90-105% | 97% | 
| EDTA plasma(n=5) | 80-95% | 87% | 
| Heparin plasma(n=5) | 83-95% | 89% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 79-97% | 96-104% | 85-93% | 89-97% | 
| EDTA plasma(n=5) | 85-94% | 87-103% | 97-105% | 79-90% | 
| Heparin plasma(n=5) | 85-102% | 78-92% | 87-98% | 95-102% |