| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.952 | 1.863 |
| 5.00 | 1.562 | 1.473 |
| 2.50 | 1.067 | 0.978 |
| 1.25 | 0.752 | 0.663 |
| 0.63 | 0.492 | 0.403 |
| 0.32 | 0.262 | 0.173 |
| 0.16 | 0.142 | 0.053 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-95% | 91% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 93-107% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-97% | 88-102% | 85-93% | 83-97% |
| EDTA plasma(n=5) | 87-98% | 85-92% | 79-96% | 89-100% |
| Heparin plasma(n=5) | 83-94% | 80-92% | 86-102% | 87-103% |