| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 30.00 | 2.327 | 2.245 |
| 15.00 | 1.649 | 1.567 |
| 7.50 | 1.078 | 0.996 |
| 3.75 | 0.754 | 0.672 |
| 1.88 | 0.586 | 0.504 |
| 0.94 | 0.357 | 0.275 |
| 0.47 | 0.175 | 0.093 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-103% | 93% |
| EDTA plasma(n=5) | 88-106% | 97% |
| Heparin plasma(n=5) | 92-114% | 103% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-114% | 87-112% | 82-117% | 84-100% |
| EDTA plasma(n=5) | 90-105% | 88-115% | 94-120% | 81-93% |
| Heparin plasma(n=5) | 83-100% | 81-108% | 91-109% | 87-100% |