| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.020 | 1.936 |
| 500.00 | 1.661 | 1.577 |
| 250.00 | 1.174 | 1.090 |
| 125.00 | 0.777 | 0.693 |
| 62.50 | 0.605 | 0.521 |
| 31.25 | 0.331 | 0.247 |
| 15.63 | 0.227 | 0.143 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-101% | 96% |
| EDTA plasma(n=5) | 91-117% | 104% |
| Heparin plasma(n=5) | 86-95% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-104% | 92-106% | 87-117% | 83-110% |
| EDTA plasma(n=5) | 86-99% | 80-106% | 83-104% | 83-101% |
| Heparin plasma(n=5) | 89-99% | 94-107% | 84-98% | 90-105% |