| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.978 | 1.879 |
| 1000.00 | 1.602 | 1.503 |
| 500.00 | 1.107 | 1.008 |
| 250.00 | 0.864 | 0.765 |
| 125.00 | 0.614 | 0.515 |
| 62.50 | 0.324 | 0.225 |
| 31.25 | 0.216 | 0.117 |
| 0.00 | 0.099 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 78-96% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-94% | 95-102% | 87-103% | 88-95% |
| EDTA plasma(n=5) | 98-107% | 90-99% | 88-97% | 85-97% |
| Heparin plasma(n=5) | 81-94% | 87-105% | 79-95% | 96-105% |